How to Measure Light Transmission of Optical Filter Glass: Methods, Specifications and Quality Control

  • Last Updated November 4, 2026
  • Written by Arnth Henriksen
Learn how to measure light transmission of optical filter glass using spectrophotometry, haze testing and quality control for OEM applications.

Table of Contents

Introduction

An optical filter glass may have a high light transmission value according to its datasheet and still perform poorly when installed in the finished device. The display appears too dark, image quality deteriorates or an optical sensor produces readings that differ from expectations.

The reason is often that transmission has been evaluated using only a single percentage value.

When determining how to measure light transmission of optical filter glass, engineers must consider more than the quantity of light passing through the material. The relevant wavelength range, measurement geometry, angle of incidence, light scattering and complete optical assembly are also essential.

For industrial OEM manufacturers, transmission is therefore more than a material characteristic. It is an important engineering parameter that influences display readability, sensor sensitivity and the reliability of optical equipment.

A dependable measurement strategy begins with the intended application and leads to clearly defined inspection procedures and tolerances for subsequent series manufacturing.

What Does Light Transmission Mean for Optical Filter Glass?

Light transmission, also called optical transmittance, describes the proportion of incident optical radiation that passes through a material.

It is normally expressed as a percentage.

If a filter glass has 80 percent transmission, it means that 80 percent of the incoming optical radiation passes through the sample and is captured under the specified measurement conditions.

The remaining radiation is reflected, absorbed or scattered into directions that are not included in the measurement.

In simplified form:

Transmission (%) = (transmitted radiant power / incident radiant power) × 100

This definition provides a useful starting point.

However, the meaning of the measurement depends on the selected wavelength range and measurement geometry.

A single transmission value does not necessarily describe the complete optical functionality of the filter glass.

Why One Percentage Value Is Not Enough

For a colour-neutral protective glass, an average transmission value in the visible spectrum may provide an initial indication of optical performance.

For technical filters, however, such a value is often insufficient.

A red filter glass, for example, may transmit red light efficiently while strongly attenuating blue and green wavelengths.

A UV-blocking filter should transmit the desired visible light while suppressing selected ultraviolet wavelengths.

An infrared filter, meanwhile, may be designed either to transmit or block IR radiation.

These different functions cannot be described adequately using a single average percentage.

The key parameter is therefore often spectral transmission, meaning transmittance as a function of wavelength.

Understanding Spectral and Visual Transmission

When measuring technical filter glass, it is important to distinguish between different transmission characteristics.

Spectral Transmission

Spectral transmission describes the proportion of radiation transmitted at a specific wavelength.

It is usually presented as a curve over a defined spectral range.

The curve shows which wavelengths the filter allows through and which it attenuates.

Spectral transmission measurements are particularly relevant for:

  • Colour filters and specialised optical filters
  • UV-blocking and IR-blocking filters
  • Bandpass and longpass filters
  • Camera windows and optical sensors
  • Filters with steep spectral transitions

A spectral measurement makes it possible to verify whether the filter provides the intended wavelength-selective performance.

Photopically Weighted Transmission

For display fronts and viewing windows intended for human observation, photopically weighted transmission can be particularly important.

This measurement weights wavelengths according to the human eye’s sensitivity under daylight viewing conditions.

The eye does not respond equally to every wavelength within the visible spectrum.

As a result, two filter glasses with different spectral characteristics may appear to have different brightness levels.

For reproducible results, the illumination spectrum, visual sensitivity function and calculation method must be specified.

Photopic transmission is therefore a useful characteristic for industrial displays, HMI front panels and illuminated indicators.

Sensor-Weighted Transmission

For cameras, photodiodes and optical sensors, transmission cannot be evaluated solely according to human visual perception.

Instead, engineers must consider the wavelengths to which the detector is sensitive.

The spectral distribution of the illumination source is equally important.

A filter glass may appear dark to the human eye while transmitting strongly in the near-infrared region.

Conversely, a visually transparent protective panel may suppress substantial portions of ultraviolet or infrared radiation.

For sensor applications, filter performance must therefore be evaluated using the actual detector sensitivity and intended illumination conditions.

How to Measure Light Transmission of Optical Filter Glass

The appropriate measurement method depends on the optical characteristic that needs to be verified.

For the development and quality control of technical optical filters, spectrophotometry and integrating-sphere measurements are particularly important.

Measuring Transmission with a Spectrophotometer

A spectrophotometer measures transmission over a specified wavelength range.

Instead of providing only a single percentage, it produces a spectral transmission curve.

This allows engineers to evaluate the optical characteristics of the material in greater detail.

Important characteristics include:

  • Transmission within the intended passband
  • Suppression of unwanted wavelengths
  • Position of spectral cutoff wavelengths
  • Steepness of transmission transitions
  • Differences between material or manufacturing batches
  • Changes following coating processes or ageing

Spectrophotometry is particularly valuable for optical sensor filters.

It can verify whether the component transmits the required wavelengths and sufficiently attenuates unwanted spectral regions.

However, the instrument’s measurement range and spectral resolution must match the application.

Measuring Direct Transmission

Direct or regular transmission measurements primarily detect light passing through the sample within a defined narrow angular range.

This method is useful for transparent optical components where the direction of transmitted light is important.

Examples include:

  • Camera windows requiring high image sharpness
  • Precision optical measurement systems
  • High-resolution display covers
  • Sensor windows with demanding optical tolerances

If the material scatters some of the light into other directions, that portion may not be captured by a measurement using a narrow acceptance angle.

This can be useful when the objective is to determine how much light remains available for precise image formation.

Measuring Total Transmission with an Integrating Sphere

When light scattering is important, an integrating sphere can be used.

An integrating sphere is designed to collect light arriving from multiple directions.

It enables measurement of total transmission, including diffuse contributions captured by the measurement arrangement.

This is particularly relevant for:

  • Matte and textured protective panels
  • Diffusers and light-scattering polymers
  • Anti-glare surfaces
  • Specialised optical coatings
  • Components designed for controlled light distribution

Comparing direct transmission with total transmission provides valuable information about scattering.

A component may have high total transmission and still reduce image sharpness.

This happens because part of the light no longer travels in the direction needed for the application.

Why Transmission and Haze Should Be Evaluated Together

For technical display fronts, transmission alone is often insufficient.

Light scattering is equally important.

One relevant parameter is haze, which describes the proportion of diffusely transmitted light under a defined test method.

Excessive haze can make a display appear less contrast-rich even when total transmission is high.

Fine text, symbols and graphical details may lose sharpness.

For a diffuser, controlled scattering may be desirable because light needs to be distributed evenly.

For camera windows and high-resolution displays, strong scattering is usually undesirable.

Transmission, haze and actual image quality should therefore be evaluated together.

The appropriate combination depends on whether the component serves as a display cover, sensor window, optical filter or diffuser.

Selecting the Correct Wavelength Range and Spectral Resolution

The measurement range must match the intended optical system.

The visible spectrum extends approximately from 380 to 780 nanometres.

However, industrial sensors frequently operate beyond this range.

Certain cameras and photodiodes, for example, detect light in the near-infrared spectrum.

Other systems require measurements in the ultraviolet region or at longer infrared wavelengths.

The wavelength range should therefore be selected according to the sensor or light source used in the finished equipment.

Spectral Resolution for Sharp Filter Transitions

Not all optical filters require the same spectral measurement resolution.

A broadband colour filter may be adequately evaluated using relatively large wavelength intervals, provided the measurement still captures its important characteristics.

Interference filters with steep transmission edges may require finer spectral resolution.

Otherwise, significant variations around the transition region could remain undetected.

This is particularly important when the filter is used with narrowband LEDs or sensors operating over a small spectral range.

Measurement resolution should therefore be matched to both the filter characteristics and the optical signal requirements.

Measurement Conditions Are Part of the Technical Specification

A transmission measurement can only be compared reliably with other results when the measurement conditions are known.

For industrial quality control, the following parameters should be documented:

  • Wavelength range used
  • Spectral resolution and measurement interval
  • Measurement geometry and instrument configuration
  • Angle of incidence
  • Sample positioning
  • Measurement aperture size
  • Surface condition and cleaning procedure
  • Reference measurement and calibration status
  • Temperature where relevant to the measurement

Without this information, results from different laboratories or suppliers may not be directly comparable.

This can create unnecessary problems during development and series manufacturing.

Critical measurement parameters should therefore form part of the product specification.

Angle of Incidence Can Change Transmission

For some optical filters, transmission depends on the angle at which light reaches the surface.

This is particularly relevant for interference-based thin-film coatings.

At oblique incidence, characteristic spectral features typically shift towards shorter wavelengths.

As a result, the passband and blocking regions may change.

A filter that achieves the required performance at normal incidence may not provide the same characteristics at larger angles.

This matters for cameras, wide-angle sensors and optical windows installed at an angle.

Measurements should therefore reflect the angles that occur in the intended device.

Where on the Filter Glass Should Measurements Be Taken?

Measurement position is also important.

For components incorporating screen printing, black border masking or localised coatings, optical properties may vary across the surface.

Measurements taken in unsuitable locations can produce misleading results.

A small measurement aperture allows individual regions to be inspected in detail.

However, it is more sensitive to positioning errors and local material variation.

A larger aperture produces more of an average over a wider area but may conceal small regions with unacceptable deviations.

Measurement locations should therefore be selected according to the component’s function and critical optical areas.

Why Finished Filter Glass Differs from Raw Material

The transmission of an untreated substrate does not necessarily represent that of the finished component.

Every additional layer and processing operation can affect optical performance.

Relevant factors include:

  • Substrate thickness and colouration
  • Anti-reflective coatings
  • Hard coatings
  • Optical adhesives and laminates
  • Additional filter films
  • Screen printing and masking
  • Electrically conductive coatings

When several optical layers are placed in series, overall transmission is often lower than the transmission of the individual layers.

Example: Transmission Through Multiple Layers

If three optical layers each transmit 90 percent of the light reaching them, the simplified total transmission is:

0.90 × 0.90 × 0.90 = 0.729

This corresponds to 72.9 percent total transmission.

The calculation assumes that the individual transmission values apply under suitable conditions and that additional optical interactions can be neglected.

In real assemblies, surface reflections, multiple internal reflections and interference effects may influence the result.

The transmission of the finished component should therefore be measured rather than relying exclusively on calculations using raw material datasheets.

Anti-Reflective Coatings and Optical Bonding

Anti-reflective coatings can reduce reflections at optical surfaces.

This may increase useful transmission and improve display readability.

However, performance depends on wavelength, angle of incidence, substrate and the complete optical stack.

Optical bonding can also provide important advantages.

In this process, the air gap between the display and cover panel is replaced by an optically clear adhesive or adhesive film.

This can reduce reflections at internal interfaces.

However, the complete optical assembly must still be evaluated.

Adhesives, coatings and substrates may affect both transmission and colour reproduction.

Temperature resistance, humidity and mechanical stresses also need to be considered.

A technically suitable solution must maintain its intended optical properties throughout expected operating conditions.

Balancing Light Transmission and EMI Shielding

For EMI shielding glass, an important trade-off exists between electromagnetic shielding and optical performance.

Electrically conductive coatings or fine metal meshes may be used to achieve the required shielding.

These structures can reduce electromagnetic interference but may also affect light transmission, colour appearance and image quality.

Metal meshes can introduce Moiré effects when their regular structure interacts with the display pixel pattern.

Maximum transmission is therefore not automatically the best design target.

If achieving higher transmission results in inadequate shielding or poor visual uniformity, the complete system may fail to meet its functional requirements.

Optical and electrical performance should therefore be specified and evaluated together.

Defining Transmission Requirements and Tolerances for Series Production

A general requirement such as “transmission at least 85 percent” is not sufficient for every industrial application.

The specification must also explain how the value is measured.

For colour or sensor filters, minimum transmission may need to be defined in particular passbands, together with maximum residual transmission in blocking regions.

For display cover panels, photopically weighted transmission combined with a maximum haze value may be more useful.

Tolerances should reflect the actual application.

Very tight limits can increase testing costs, rejection rates and manufacturing complexity.

Excessively broad tolerances can lead to visible colour differences, uneven display brightness or inadequate separation of sensor signals.

The key question is which variations the complete system can tolerate without losing the required functionality.

Development Testing, First Article Inspection and Production Control

An effective inspection strategy should distinguish between different stages of manufacturing.

During development, detailed spectral measurements help engineers select substrates, coatings and optical constructions.

During first article inspection, the finished component is evaluated against the agreed specifications.

During series production, more targeted, risk-based inspection procedures may be appropriate.

It is not always necessary to perform a complete spectral scan on every manufactured component.

However, simplified procedures require evidence that the manufacturing process is sufficiently capable and that the inspection method correlates reliably with the agreed reference measurement.

This approach can combine dependable quality control with commercially efficient manufacturing.

Common Errors When Measuring Light Transmission

Many measurement problems arise not from the instrument itself, but from sample condition or handling.

Contamination, Fingerprints and Protective Films

Fingerprints, dust, cleaning residues and protective films can alter the measured transmission.

Even small amounts of contamination may influence results for highly transparent substrates.

Samples should therefore be cleaned and handled according to defined procedures.

Direct contact with critical optical surfaces should be avoided wherever practical.

Incorrect Mounting and Mechanical Stress

A sample that is mounted incorrectly or mechanically stressed during measurement may exhibit optical behaviour that differs from its intended condition.

This can be particularly relevant for certain polymer components and sensitive optical structures.

The measurement fixture should therefore avoid introducing unnecessary deformation.

Confusing High Transmission with Good Readability

A common misconception is that a display always becomes easier to read when the cover glass transmits more light.

This is not necessarily true.

A darker front panel may, under certain lighting conditions, reduce distracting reflections and unwanted ambient light.

This can improve perceived contrast even though less light from the display passes through the panel.

Conversely, a highly transparent cover panel may provide poor readability if its surface reflects too much ambient light.

Transmission, reflectance, display brightness, ambient illumination and viewing angle must therefore be evaluated together.

Select Sensor Filters Based on Signal Quality

For optical sensor applications, maximum light transmission is not always the main objective.

A filter may transmit a high proportion of the desired light while also allowing unwanted radiation to reach the detector.

This can reduce measurement quality.

What matters is often the relationship between the useful sensor signal and unwanted optical contributions.

The filter’s spectral characteristics must therefore match the sensor’s sensitivity and the actual illumination spectrum.

Angle of incidence, stray light, mechanical integration and temperature changes may also affect the result.

The goal should be stable and repeatable sensor performance rather than the highest possible transmission at one particular wavelength.

Measure and Specify Optical Filter Glass Transmission Reliably

Reliable light transmission measurement requires more than a single percentage value.

Spectral transmission, photopically weighted transmission, direct and total transmission, and haze provide different information about the optical performance of a component.

The most important characteristics depend on the application.

For industrial displays, readability, contrast and optical uniformity are often central.

For sensors, defined passbands, blocking regions and actual signal quality may be more important.

PSC develops technical optical filter and cover glass solutions with consideration for material selection, coatings, geometry and industrial manufacturability.

A measurement strategy established early in development creates a stronger foundation for representative prototypes and consistent quality control during series production.

When optical filter glass transmission is evaluated using clearly defined wavelength ranges, measurement conditions and functional tolerances, the result becomes more than a datasheet value. It provides a reliable basis for developing industrial optical equipment and consistently manufactured components.

Picture of Arnth Henriksen

Arnth Henriksen

Arnth Henriksen is an experienced leader with an international background and extensive expertise in sales, marketing, and management. He has worked across multiple industries and has gained solid experience in driving growth and optimizing business strategies. His ability to lead teams and navigate complex international markets has been a key factor in his success. With his broad experience and analytical approach, he has achieved impressive results in both established and emerging markets.
Read more articles by Arnth Henriksen
LinkedIn: LinkedIn
Email: ah@psc.dk

WE ARE READY TO HELP YOU WITH A UNIQUE OPTICAL FILTER SOLUTION

We are ready to help you with a unique optical filter solution.

PSC / CLAREX

Company

About Us

Our Process

Articles

Documentation

Solutions

Display Cover Glass

EMI Shielding

Optical Filters

Custom Solutions

Support

Contact Us

Privacy Policy

Terms of Service

Newsletter

Stay updated with our latest innovations and solutions.

Subscribe to our newsletter

Find us
  • PSC A/S
  • Industrivej 13
  • DK-9700 Brønderslev
  • Denmark
  • CVR: 27852610​
Contact us
  • +45 98 82 29 99
  • contact@psc.dk
Linkedin

Newsletter

Stay updated with our latest innovations and solutions.

Subscribe to our newsletter

© 2024: PSC – MANUFACTURER OF OPTICAL FILTERS AND DISPLAY WINDOW SOLUTIONS

PSC assumes no responsibility for errors or omissions in the content of this site.

Information such as product specifications and technical data are not legally binding and are subject to change.

  • Privacy and Cookie Policy
  • Terms of sale & delivery
Main Menu
  • Home
  • Solutions
    • Display Cover Glass
    • Optical Filters
    • EMI Shielding
  • Base Materials
    • Acrylic (PMMA)
    • Polycarbonate (PC)
    • Glass
  • Product Catalog
  • Cases
  • Company
    • About Us
    • Our Process
  • Resources
    • Articles
    • Documentation
Get in touch